The use and maintenance of sterilization pots
Source:本站 Time:2024/4/18 14:42:18 frequency:
The use of sterilization pots should strictly follow the safety operating procedures. The safety operating procedures for sterilization pots are as follows:
(1) The operating process indicators, maximum working pressure, and maximum working temperature of the sterilization pot& Nbsp;
(2) Operation methods, procedures, and precautions for sterilization pots& Nbsp;
(3) During production and operation, key inspections should be carried out on items and parts, as well as any possible abnormal phenomena and preventive measures that may occur during operation& Nbsp;
(4) Maintenance methods during equipment shutdown& Nbsp;
Operators using sterilization pots should undergo training and pass exams, and strictly adhere to safety operating procedures and job responsibility systems; Any abnormal phenomena should be dealt with promptly.
The sterilization pot is equipped with safety valves, pressure gauges, thermometers and other accessories, which should be kept safe, complete, sensitive, and reliable. Maintenance and regular verification should be carried out during use. The opening and closing pressure of the safety valve of this equipment is equal to the design pressure, and it should be kept sensitive and reliable, and should be prevented from being adjusted arbitrarily. The accuracy level of pressure gauges and thermometers is both 1.5, and the difference within the allowable tolerance is normal. Pressure gauges without lead seals should not be used. If the pressure gauge indicates malfunction, the scale is unclear, the dial is broken, the pointer does not return to zero after pressure relief, and the lead seal is damaged, it should be replaced immediately. Thermometers should be regularly inspected and calibrated with a standard thermometer before use. They should be calibrated once a year thereafter. Any mercury column that breaks or deviates from the standard temperature by more than 0.5 ℃ should be sent for repair or replacement& Nbsp;
The sterilization pot should be inspected regularly, with external inspections conducted at least once every six months and comprehensive inspections conducted at least once a year. The preparation work and inspection items before the inspection should be carried out in accordance with the requirements; Regulations; According to relevant regulations, the inspection report shall be archived and filed& Nbsp;
During use, lubricating grease and oil should be regularly applied to sliding and rolling contact areas such as inclined self-locking wedges, rollers, pot ears, and swivel flanges to ensure that the friction surface is in a lubricated state.
When the equipment is out of use, the inner and outer surfaces of the sterilization pot and lid should be cleaned thoroughly. The exposed processing surface is coated with anti rust paint, and the safety inspection and control instruments are covered and protected.